Gruber, V; Morakeas, S; Hinder, M; Drevhammar, T; Dronavalli, M; Tracy, MB.
Leak during term neonatal CPAP stabilisation: does resuscitation device design affect delivered PEEP - a bench study.
BMJ Paediatr Open. 2025; 9(1): Doi: 10.1136/bmjpo-2025-003679 [OPEN ACCESS]
Web of Science PubMed PUBMED Central FullText FullText_MUG

 

Führende Autor*innen der Med Uni Graz
Gruber Viktoria
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
BACKGROUND: Mask leak is common during newborn resuscitation, impacting delivered ventilation. Less is known about the effect of leak during assisted spontaneous breathing with continuous positive airway pressure (CPAP). AIM: Compare CPAP performance in the presence of leak of two T-piece resuscitation devices, with differing design-imposed resistances: high resistance Neopuff and low resistance rPAP. METHODS: Devices were tested on a term dynamic lung model (compliance 1.0 mL/cmH2O, tidal volume (VT) 23 mL) at three incremental leak levels and set positive end expiratory pressure (PEEP) 5 cmH2O, 7 cmH2O and 9 cmH2O. Continuous leaks were generated by differing-length tubes open to atmosphere. Measured parameters were mean pressure, fluctuations around set pressure (∆P), VT and leak flow. RESULTS: 2437 breaths were analysed (1216 rPAP, 1221 Neopuff). Leak reduced PEEP, with the largest reduction at the highest leak and 9 cmH2O set PEEP (Neopuff 2.6 cmH2O vs rPAP 7.0 cmH2O). Higher delivered PEEP was associated with higher leak flows (Neopuff 4.4 L/min vs rPAP 8.2 L/min, 9 cmH2O set PEEP). VT reduced with Neopuff compared with rPAP (Neopuff 18 mL vs rPAP 23.2 mL, no leak, 9 cmH2O set PEEP) and was affected by delivered PEEP. CONCLUSION: The delivered support differed between devices in the presence of leaks. rPAP maintained pressures closer to the set PEEP value at all leak levels, and higher leak flows were seen with the maintained distending pressure. Neopuff's higher resistance led to reductions in VT that were more pronounced at low leak levels.
Find related publications in this database (using NLM MeSH Indexing)
Continuous Positive Airway Pressure - instrumentation
Equipment Design - administration & dosage
Humans - administration & dosage
Infant, Newborn - administration & dosage
Positive-Pressure Respiration - instrumentation
Resuscitation - instrumentation
Tidal Volume - administration & dosage
Equipment Failure - administration & dosage
Masks - administration & dosage