Winkler, P; Zurl, B; Guss, H; Kindl, P; Stuecklschweiger, G.
Performance analysis of a film dosimetric quality assurance procedure for IMRT with regard to the employment of quantitative evaluation methods.
PHYS MED BIOL. 2005; 50(4): 643-654. Doi: 10.1088/0031-9155/50/4/006
Web of Science PubMed FullText FullText_MUG

 

Führende Autor*innen der Med Uni Graz
Winkler Peter
Co-Autor*innen der Med Uni Graz
Zurl Brigitte
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
A system for dosimetric verification of intensity-modulated radiotherapy (IMRT) treatment plans using absolute calibrated radiographic films is presented. At our institution this verification procedure is performed for all IMRT treatment plans prior to patient irradiation. Therefore clinical treatment plans are transferred to a phantom and recalculated. Composite treatment plans are irradiated to a single film. Film density to absolute dose conversion is performed automatically based on a single calibration film. A software application encompassing film calibration, 2D registration of measurement and calculated distributions, image fusion, and a number of visual and quantitative evaluation utilities was developed. The main topic of this paper is a performance analysis for this quality assurance procedure, with regard to the specification of tolerance levels for quantitative evaluations. Spatial and dosimetric precision and accuracy were determined for the entire procedure, comprising all possible sources of error. The overall dosimetric and spatial measurement uncertainties obtained thereby were 1.9% and 0.8 mm respectively. Based on these results, we specified 5% dose difference and 3 mm distance-to-agreement as our tolerance levels for patient-specific quality assurance for IMRT treatments.
Find related publications in this database (using NLM MeSH Indexing)
Benchmarking - methods Benchmarking - standards
Equipment Failure Analysis - methods Equipment Failure Analysis - standards
Film Dosimetry - instrumentation Film Dosimetry - methods Film Dosimetry - standards
Quality Assurance, Health Care - methods Quality Assurance, Health Care - standards
Radiotherapy Dosage -
Radiotherapy Planning, Computer-Assisted - instrumentation Radiotherapy Planning, Computer-Assisted - methods
Radiotherapy, Conformal - methods Radiotherapy, Conformal - standards
Reproducibility of Results -
Sensitivity and Specificity -