Disparity of time-contrast curves generated by various types of power injectors used in magnetic resonance imaging.
Sci Rep. 2020; 10(1): 19568 Doi: 10.1038/s41598-020-76536-x [OPEN ACCESS]
Web of Science PubMed PUBMED Central FullText FullText_MUG

 

Co-authors Med Uni Graz
Klickovic Uros
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
The profiles of time-contrast (TC) -curves from popular MRI injectors derived at the injection site of the attached tube-line system were compared. Variations of TC-profiles were previously reported to potentially influence image quality in time critical MRI measurements. TC-curves from five injectors obtained during commonly used injection protocols were assessed according to representative quality criteria: (1) correlation strength between a fitted boxcar function and the TC-curve (cBCF) and (2) difference between true and expected injection time (dBIT). Additionally, the impact from technical injector properties: pump type, line volume, maximum injection power and type of contrast medium (CM) on the TC-profiles was evaluated. Injectors using a piston-syrinx (PS) mechanism for CM-injection performed significantly better than those working with a peristaltic roller pump (RP) technique. Besides injection mechanism, line filling volume showed a strong influence on the final TC-curves, where larger filling volumes induced worse cBCF- and dBIT-results. Therefore, to achieve an optimal bolus in clinical MRI use of a PS-injector seems recommendable. Besides their pump mechanism, RP-injectors appeared additionally hampered by their high volume line systems, pointing out an unfavourable coinicidence of these technical features in RP-injectors. This should be considered, particularly, in comparative or time-critical MRI-studies.

© Med Uni GrazImprint