Hausegger, KA; Tiesenhausen, K; Tauss, J; Karaic, R; Klein, GE; Aman, W; Koch, G.
Endoluminal therapy of infrarenal aortic aneurysms with the Talent system--initial experiences with a new endoprosthesis
Rofo. 1998; 169(6):633-638 Doi: 10.1055/s-2007-1015354
Web of Science PubMed FullText FullText_MUG

 

Führende Autor*innen der Med Uni Graz
Hausegger Klaus
Co-Autor*innen der Med Uni Graz
Klein Guenther
Tauss Josef
Tiesenhausen Kurt
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
PURPOSE: To report the early clinical experience with the Talent stent-graft system in the treatment of infrarenal aortic aneurysms. MATERIALS AND METHODS: The study group comprised 15 males aged from 51-79 years. A bifurcated Talent stent-graft was implanted in 9 patients with type B or C aneurysms, a tube graft in 6 patients with a type A aneurysm. The stent-grafts were introduced via a surgical femoral cut-down, for a bifurcated graft a bilateral surgical access was needed. All procedures were performed as teamwork by radiologists and vascular surgeons under general anaesthesia in the angiography suite. RESULTS: Primary and secondary technical successes were 80 and 86%, respectively. Technical failures were due to access site problems (n = 1; elective open repair), malpositioning of the prosthesis (n = 1; immediate conversion to open surgery), and primary distal leakage (n = 1; elective occlusion by means of distal stent-graft extension). One patient died 15 days after an uneventful stent-graft insertion (30-day mortality 6%). No complications were observed during a mean follow-up period of 7.8 months. The aneurysmal diameter decreased in 6 patients and remained unchanged in the others. CONCLUSIONS: The Talent system revealed satisfactory early results which are comparable to competitive stent-graft systems. An advantage of this stent-graft device is the availability in a wide range of dimensions. A major disadvantage is the large introducer system (french size 22-27).
Find related publications in this database (using NLM MeSH Indexing)
Aged -
Angioplasty, Balloon - instrumentation
Aortic Aneurysm, Abdominal - mortality
Aortography - mortality
Blood Vessel Prosthesis - mortality
Coated Materials, Biocompatible - mortality
Female - mortality
Follow-Up Studies - mortality
Humans - mortality
Male - mortality
Middle Aged - mortality
Prosthesis Design - mortality
Stents - mortality
Survival Rate - mortality
Tomography, X-Ray Computed - mortality

Find related publications in this database (Keywords)
Abdominal Aortic Aneurysm
Therapy
Stent-Graft
Endoluminal Treatment
© Med Uni Graz Impressum