Tiran, A; Tiran, B; Hojas, S; Kostner, GM; Wilders-Truschnig, MM.
Immunoquantification of lipoprotein(a): comparison of nephelometry with electroimmunodiffusion.
J Clin Lab Anal. 1993; 7(5):256-262 Doi: 10.1002/jcla.1860070504
Web of Science PubMed FullText FullText_MUG

 

Leading authors Med Uni Graz
Tiran Andreas
Co-authors Med Uni Graz
Kostner Gerhard
Tiran Beate
Truschnig-Wilders Martini
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
A new fully automated nephelometric immunoassay for lipoprotein(a) quantification in human serum was evaluated using the Behring Nephelometer Analyzer. The assay exhibited a good linearity in the concentration range of 110-1,770 mg/l; at higher concentrations, samples were automatically diluted by a factor of 4. The method is simple, robust, and shows an excellent stability of the calibration curve over several weeks. Intra-assay and day-to-day coefficients of variation were 2% and 4.5%, respectively. The method correlated well with electroimmunodiffusion (r = 0.977; n = 123; P = 0.0001). Unspecific turbidity as expressed by an elevated blank value occurred in 3% of all freshly measured samples (n = 392). Storage of the samples for 1 week at 4 degrees C had no significant influence on the results. Frozen sera, on the other hand, cannot be assayed by this method. We believe that this assay is well suited for use in clinical routine work.
Find related publications in this database (using NLM MeSH Indexing)
Calibration -
Humans -
Immune Sera - immunology
Immunoassay - methods
Immunodiffusion - methods
Lipoprotein(a) - analysis
Nephelometry and Turbidimetry - analysis
Sensitivity and Specificity - analysis

Find related publications in this database (Keywords)
Lipoprotein(A) Quantification
Behring Nephelometer Analyzer