Eckstein, K; Bachrata, B; Hangel, G; Widhalm, G; Enzinger, C; Barth, M; Trattnig, S; Robinson, SD.
Improved susceptibility weighted imaging at ultra-high field using bipolar multi-echo acquisition and optimized image processing: CLEAR-SWI.
Neuroimage. 2021; 237:118175 Doi: 10.1016/j.neuroimage.2021.118175 [OPEN ACCESS]
Web of Science PubMed PUBMED Central FullText FullText_MUG

 

Führende Autor*innen der Med Uni Graz
Robinson Simon Daniel
Co-Autor*innen der Med Uni Graz
Enzinger Christian
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
PURPOSE: Susceptibility Weighted Imaging (SWI) has become established in the clinical investigation of stroke, microbleeds, tumor vascularization, calcification and iron deposition, but suffers from a number of shortcomings and artefacts. The goal of this study was to reduce the sensitivity of SWI to strong B1 and B0 inhomogeneities at ultra-high field to generate homogeneous images with increased contrast and free of common artefacts. All steps in SWI processing have been addressed - coil combination, phase unwrapping, image combination over echoes, phase filtering and homogeneity correction - and applied to an efficient bipolar multi-echo acquisition to substantially improve the quality of SWI. PRINCIPAL RESULTS: Our findings regarding the optimal individual processing steps lead us to propose a Contrast-weighted, Laplace-unwrapped, bipolar multi-Echo, ASPIRE-combined, homogeneous, improved Resolution SWI, or CLEAR-SWI. CLEAR-SWI was compared to two other multi-echo SWI methods and standard, single-echo SWI with the same acquisition time at 7 T in 10 healthy volunteers and with single-echo SWI in 13 patients with brain tumors. CLEAR-SWI had improved contrast-to-noise and homogeneity, reduced signal dropout and was not compromised by the artefacts which affected standard SWI in 10 out of 13 cases close to tumors (as assessed by expert raters), as well as generating T2* maps and phase images which can be used for Quantitative Susceptibility Mapping. In a comparison with other multi-echo SWI methods, CLEAR-SWI had the fewest artefacts, highest SNR and generally higher contrast-to-noise. MAJOR CONCLUSIONS: CLEAR-SWI eliminates the artefacts common in standard, single-echo SWI, reduces signal dropouts and improves image homogeneity and contrast-to-noise. Applied clinically, in a study of brain tumor patients, CLEAR-SWI was free of the artefacts which affected standard, single-echo SWI.

Find related publications in this database (Keywords)
SWI
CLEAR-SWI
Ultra-high Field
Multi-echo
Brain tumor
T2*