Friesenbichler, J; Schwarzkopf, R; Sadoghi, P; Marwin, SE; Glehr, M; Maurer-Ertl, W; Leithner, A.
Failure rate of a rotating hinge knee design due to yoke fracture of the hinged tibial insert: a retrospective data analysis and review of the literature.
Int Orthop. 2012; 36(5):993-998 [OPEN ACCESS]
Web of Science PubMed PUBMED Central FullText FullText_MUG

 

Autor/innen der Med Uni Graz:
Friesenbichler Jörg
Glehr Mathias
Leithner Andreas
Maurer-Ertl Werner
Sadoghi Patrick
Altmetrics:

Dimensions Citations:

Plum Analytics:
Number of Figures: 4
| | | |
Abstract:
Rotating hinge knee prostheses are known to provide inherent stability. Yoke fractures of the hinged tibial insert of modern generation rotating hinge devices are a matter of continued concern. The aim of this study was to describe incidence and management of yoke fracture of the LPS (TM) hinged tibial insert. Retrospective data analysis of two institutions identified 40 patients with a LPS (TM) total knee arthroplasty. Implant survival and prosthetic complications was calculated according to Kaplan-Meier. Out of the group of 40 patients, four fractures of the metal yoke occurred in four cases (failure rate: 10%). Furthermore, a second fracture occurred in two patients. The overall revision-free prosthetic survival was 57% at 38 months, while prosthetic survival until yoke fracture was 86% at 38 months. Handling yoke fractures as mechanical complication includes replacing the hinged insert, stabilization of the joint and joint line height preservation in order to decrease the cantilever effect at the insert-base plate interface.
Find related publications in this database (using NLM MeSH Indexing)
Adolescent -
Adult -
Aged -
Aged, 80 and over -
Arthroplasty, Replacement, Knee -
Equipment Failure Analysis -
Female -
Humans -
Kaplan-Meier Estimate -
Knee Joint - surgery
Knee Prosthesis -
Male -
Middle Aged -
Postoperative Complications -
Prosthesis Design -
Prosthesis Failure -
Retrospective Studies -
Rotation -
Tibia - surgery
Young Adult -