Manninger, M; Odeneg, T; Fruhwald, F; Brussee, H; Scherr, D.
Oversensing of the wearable cardioverter defibrillator during bipolar ventricular stimulation.
Wien Klin Wochenschr. 2017; 129(23-24):910-912
Doi: 10.1007/s00508-017-1290-z
[OPEN ACCESS]
Web of Science
PubMed
FullText
FullText_MUG
- Führende Autor*innen der Med Uni Graz
-
Manninger-Wünscher Martin
- Co-Autor*innen der Med Uni Graz
-
Brussee Helmut
-
Fruhwald Friedrich
-
Scherr Daniel
- Altmetrics:
- Dimensions Citations:
- Plum Analytics:
- Scite (citation analytics):
- Abstract:
-
The wearable cardioverter defibrillator (WCD) is a temporary treatment option for patients with potentially reversible risk of sudden cardiac death. This case demonstrates a pitfall during WCD usage in a pacemaker-dependent patient as well as a possible solution allowing continuation of WCD therapy. Bipolar stimulation may lead to double counting of the WCD detection algorithm resulting in false alarm or inappropriate therapy.
- Find related publications in this database (Keywords)
-
Pacing
-
Myocarditis
-
False alarm
-
Misdetection
-
Telemedicine