Comparison of frequency-domain and continuous-wave near-infrared spectroscopy devices during the immediate transition.
BMC Pediatr. 2020; 20(1):94-94 Doi: 10.1186/s12887-020-1987-4 [OPEN ACCESS]
Web of Science PubMed PUBMED Central FullText FullText_MUG

 

Co-Autor*innen der Med Uni Graz
Pichler Gerhard
Urlesberger Berndt
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
Non-invasive monitoring of cerebral tissue oxygen saturation (rcSO2) during transition is of growing interest. Different near-infrared spectroscopy (NIRS) techniques have been developed to measure rcSO2. We compared rcSO2 values during the immediate transition in preterm neonates measured with frequency-domain NIRS (FD-NIRS) with those measured with continuous-wave NIRS (CW-NIRS) devices in prospective observational studies. We compared rcSO2 values measured with an FD-NIRS device during the first 15 min after birth in neonates with a gestational age ≥ 30 weeks but < 37 weeks born at the Erasmus MC- Sophia Children's Hospital, Rotterdam, the Netherlands, with similar values measured with a CW-NIRS device in neonates born at the Medical University of Graz, Austria. Mixed models were used to adjust for repeated rcSO2 measurements, with fixed effects for time (non-linear), device, respiratory support and the interaction of device and respiratory support with time. Additionally, parameters such as total haemoglobin concentration and oxygenated and deoxygenated haemoglobin concentrations measured by FD-NIRS were analysed. Thirty-eight FD-NIRS measurements were compared with 58 CW-NIRS measurements. The FD-NIRS rcSO2 values were consistently higher than the CW-NIRS rcSO2 values in the first 12 min, irrespective of respiratory support. After adjustment for respiratory support, the time-dependent trend in rcSO2 differed significantly between techniques (p < 0.01). As cerebral saturation measured with the FD-NIRS device differed significantly from that measured with the CW-NIRS device, differences in absolute values need to be interpreted with care. Although FD-NIRS devices have technical advantages over CW-NIRS devices, FD-NIRS devices may overestimate true cerebral oxygenation and their benefits might not outweigh the usability of the more clinically viable CW-NIRS devices.

Find related publications in this database (Keywords)
Near-infrared spectroscopy
Continuous-wave
Frequency-domain
Transition