Morris, N; Pichler, G; Pocivalnik, M; Brandner, A; Müller, W; Urlesberger, B.
Cerebral regional oxygen saturation (crSO2): are different sensors comparable?
Paediatr Anaesth. 2012; 22(11):1132-1134
Doi: 10.1111/j.1460-9592.2012.03895.x
Web of Science
PubMed
FullText
FullText_MUG
- Führende Autor*innen der Med Uni Graz
-
Morris Nicholas Mark
-
Pichler Gerhard
- Co-Autor*innen der Med Uni Graz
-
Brandner Anna
-
Müller Wilhelm
-
Ribitsch Mirjam
-
Urlesberger Berndt
- Altmetrics:
- Dimensions Citations:
- Plum Analytics:
- Scite (citation analytics):
- Abstract:
-
For measurement of cerebral regional oxygen saturation (crSO2) in neonates, one of the frequently used near-infrared spectroscopy devices is the INVOS 5100 with either the neonatal or pediatric sensor. Measurements between adult and pediatric sensors use different algorithms and differ by 10%. There are no published data comparing neonatal and pediatric sensors.
Aim of this study was to compare the neonatal sensor with the pediatric sensor of the INVOS 5100 device in a mixed cohort of neonates to recognize possible differences and enable interpretation of data.
In 40 neonates, crSO2 was measured in identical positions using both the neonatal and the pediatric sensor under identical conditions. Each sensor was reapplied five times to calculate mean values and standard deviations.
Cerebral regional oxygen saturation measurements using the neonatal sensor (mean, 76.1 ± 7.9) and crSO2 measurements using the pediatric sensor (mean, 76.4 ± 8.1) were not different. Reproducibility was not significantly different with similar means of standard deviations (neonatal, 3.7 ± 1.6; pediatric, 3.1 ± 1.3).
In a mixed cohort of neonates, crSO2 measurements were identical when using the INVOS 5100 neonatal sensor and the pediatric sensor. We conclude that studies using the INVOS 5100 neonatal or pediatric sensor are well comparable.
© 2012 Blackwell Publishing Ltd.
- Find related publications in this database (using NLM MeSH Indexing)
-
Brain - metabolism
-
Cohort Studies -
-
Female -
-
Humans -
-
Infant, Newborn -
-
Male -
-
Oxygen - metabolism
-
Prospective Studies -
-
Reproducibility of Results -
-
Spectroscopy, Near-Infrared - instrumentation
- Find related publications in this database (Keywords)
-
near-infrared spectroscopy
-
cerebral regional oxygen saturation
-
sensors
-
comparison
-
neonates